Yokogawa YTA610 Temperature Transmitter Field Mount
The YTA610 is the highly accurate temperature transmitter that accepts Thermocouple, RTD, ohms or DC millivolts inputs and converts it to a 4 to 20 mA DC or Fieldbus signal for transmission.
The YTA610 supports HART® and FOUNDATION fieldbusTM communication protocols.HART types are certified as complying with SIL 2 for safety requirement.
The YTA610 provides high accuracy, stability, and reliability. The housing structure adopts reliable dual-compartment structure. The sensor input can be choice of single or dual inputs. This input can accept the RTD, thermocouple, ohm, or DC millivolt. It converts the sensor input to a 4 to 20 mA DC analog signal or Fieldbus digital signal. HART® 7 Protocol or FOUNDATION™ Fieldbus ITK 6 version are available. The dual input can accept two sensors calculation of the differential or average value and the sensor backup function. Sensor backup functions for automatically switches-over from the primary to the backup upon sensor failure.
By adoption of local parameter switch on the Indicator, it becomes easy to do the on-site setting work.
YTA610 features include:
- Dual-compartment field-mount housing
- Choice of HART® 7, or FOUNDATION™ Fieldbus ITK 6 Protocol
- Local Parameter Setting function (Indicator model)
- Basic diagnosis
STANDARD SPECIFICATIONS
PERFORMANCE SPECIFICATIONS
Accuracy
HART communication type:
A/D accuracy/span + D/A accuracy
(See Table 1 on page 4.)
Fieldbus communication type:
A/D accuracy (See Table 1 on page 4.)
Cold Junction Compensation Accuracy
± 0.5°C (± 0.9 °F) for T/C only
Ambient Temperature Effect (per 10°C change)
±0.1% or ±(Temperature coefficient/span), whichever
is greater. (See Table 2 on page 5.)
Stability
RTD: ±0.1% of reading or ±0.1°C per 2 years,
whichever is greater at 23±2°C.
T/C: ±0.1% of reading or ±0.1°C per year, whichever
is greater at 23±2°C.
5 Year Stability
RTD: ±0.25% of reading or ±0.25°C,
whichever is greater at 23±2°C.
T/C: ±0.5% of reading or ±0.5°C,
whichever is greater at 23±2°C.
Vibration Effect
The YTA610 is tested to the following specifications
with no effect on performance per IEC 60770-1
10 to 60 Hz 0.21 mm peak displacement
60 to 2000 Hz 3g
Power Supply Effect
±0.005% of calibrated span per volt